(Go back one page, or back to the Introduction).
Four steps must be taken to assure proper compensation. These steps should
be taken with every experiment where the cell types differ, where the reagents
differ, or where any instrument settings change--daily compensation is thus
probably not frequent enough!
(1) The compensation tube must consist of cells that are unstained as well
as cells that are singly-stained with the fluorescent probes. The stained
(positive) cells must have the same autofluorescence (when they are unstained)
as do the unstained (negative) cells in the compensation tube (e.g., all
are lymphocytes).
(2) The PMT voltages must be set high enough guarantee that the negative
population is off the axis in every channel.
(3) An analysis gate is set so that only cells with identical autofluorescence
characteristics are viewed (e.g., a lymphocyte gate). An analysis gate is
also set to include all of the negative cells and all of the positive cells.
(4) The centers of the positive and negative cell populations are aligned
by matching the median fluorescences.
Finish off!